Ivoclar Programat P90 Manual Fixed Info
Critical Note: You must insert the new battery within of removing the old one, or all stored program data will be lost.
The heating muffle should be replaced if heating becomes inadequate or if the furnace displays Er23 or Er24 . Avoid touching the quartz tube or heating wire with bare fingers to prevent premature damage from grease contamination. ivoclar programat p90 manual fixed
If the vacuum cannot be established, check the sealing rim (69) and sealing ring (35) for debris or damage. Clean these components regularly with a dry or slightly moist cloth; never use solvents. Troubleshooting Common Error Codes Critical Note: You must insert the new battery
For vacuum programs, ensure the red light spot appears after the appropriate digits in the display indicator to confirm the vacuum switch-on/off points are active. Ivoclar Programat p90 Operating Instructions | PDF - Scribd If the vacuum cannot be established, check the